16 Jun
2014
16 Jun
'14
5:39 p.m.
That is different from the CCP4 lib. Thanks for pointing this out! $ grep C2D $CCP4/lib/data/monomers/h/HEM.cif | grep C3D ~ HEM C2D C3D double 1.352 0.020 ~ eab On 06/16/2014 01:13 PM, Yarrow Madrona wrote:
Hello,
After validating a PDB for deposition I received a message that the HEME ligand C2D-C3D distance was longer than ideal. It was 1.52 angstroms, while it was 1.35 for C2A-C3A, C2B-C3B and C2C-C3C.
The ideal bond length is listed as 1.544 vs. 1.337 for the other bonds in $PHENIX/chem_data/mon_lib/h/HEM.cif
Is there a reason for this? If not, I am inclined to edit it.
-Yarrow
_______________________________________________ phenixbb mailing list [email protected] http://phenix-online.org/mailman/listinfo/phenixbb