Hi Qun, - I agree with Ralf: the version you use is really old and I would suggest using the latest available that you can download from this web page: http://www.phenix-online.org/download/ - When you run joint XN refinement, you need to explicitly specify which data is neutron and which is X-ray (there is no (reliable) way for phenix.refine to figure this out automatically). So, I would correct your command below as following and run again: phenix.refine xray_data.file_name=dhfr-mtx-rt-xray.hkl neutron_data.file_name=newdhfr_un_change.hkl xn_easy_080326a.pdb input.xray_data.labels=FOBSx input.neutron_data.labels=FOBSn xray_data.r_free_flags.generate=True neutron_data.r_free_flags.generate=True refine_default.params The fraction of your file "refine_default.params" looks good to me. The choice for scattering dictionary is automatically adjusted given the input data. I agree it is a bit confusing the way it's reported in parameters file, I'm making a note of this right now and I will make it clearer in future. A few points to keep in mind doing joint XN refinement: - you can automatically add H atoms as well as exchangeable H/D using phenix.ready_set. Example: phenix.ready_set model.pdb - when doing joint XN refinement, try using "optimize_wxc=true optimize_wxu=true". This might be essential to get a good result. - also, I would recommend running separate X-ray and Neutron refinements to compare with. Please let me know if you have any questions or problems with this; I will be happy to help with any issues. Cheers, Pavel. On 1/1/2009 9:15 PM, crystallogrphy wrote:
Deall all:
I have a question about xray and neutron joint refinement.
When I start phenix joint refinement like this, phenix gave the wrong message: phenix.refine dhfr-mtx-rt-xray.hkl neutron_data.file_name= newdhfr_un_change.hkl xn_easy_080326a.pdb input.xray_data.labels=FOBSx input.neutron_data.labels=FOBSn xray_data.r_free_flags.generate=True neutron_data.r_free_flags.generate=True refine_default.params
===================== joint neutron/xray refinement start =====================
Traceback (most recent call last): call_back_after_monitor_collect=call_back_after_monitor_collect) File "/usr/local/phenix-1.3-rc4/phenix/phenix/refinement/driver.py", line 1031, in run call_back_after_monitor_collect = call_back_after_monitor_collect) File "/usr/local/phenix-1.3-rc4/phenix/phenix/refinement/strategies.py", line 184, in refinement_machine * assert not neutron_refinement AssertionError*
I guess in my default parameters, something is wrong, particularly the bold word I highlighted. But I do not know how to fix. Does any one have a clue?\ Thanks and Happy New Year!
main { bulk_solvent_and_scale = True simulated_annealing = False ordered_solvent = False ncs = False ias = False number_of_macro_cycles = 3 max_number_of_iterations = 25 use_form_factor_weights = False tan_u_iso = False use_convergence_test = False target = *ml mlhl ml_sad ls min_number_of_test_set_reflections_for_max_likelihood_target = 50 max_number_of_resolution_bins = 30 reference_xray_structure = None use_experimental_phases = None compute_optimal_errors = False random_seed = 2679941 *scattering_table = wk1995 it1992 *n_gaussian neutron* use_normalized_geometry_target = True target_weights_only = False use_f_model_scaled = False max_d_min = 0.25 fake_f_obs = False
-- Qun Wan Department of Pharmacology, Case Western Reserve University 10900 Euclid Avenue, Cleveland, OH 44106 email: [email protected] mailto:[email protected] lab phone: 216-368-3337 ------------------------------------------------------------------------